Microsep® - Model 1140 Separometer - Use Nitrile Gloves

Microsep® - Model 1140 Separometer - Use Nitrile Gloves

From The Technical Insights Blog

Posted by: Emcee Electronics on Friday, February 21, 2014
Category: Tech Tips
Tags:

Our Model 1140 Micro-Separometer is routinely used to perform the tests covered by  ASTM methods D3948, D4860, D7224, and D7261. We caution our customers to please  use nitrile gloves rather than latex when doing any of these tests. We have found that the  use of latex gloves may cause you to have faulty results because of the buildup of static electricity inherent in the use of latex. Additionally, nitrile rubber offers more resistance to petroleum products.